Narishige Pc10 Manual New |best| (2025)

Controls the electrical current sent to the platinum coil during the first and second pull stages. Higher numbers increase the temperature.

: Adjust the position plate and set the "No.1 Heater" level. The capillary will pull down to a thin, tapered state. Automated Reset

This public link is valid for 7 days and shares a thread, including any personal information you added. This link or copies made by others cannot be deleted. If you share with third parties, their policies apply. Can’t copy the link right now. Try again later.

Removable metal disks located on the bottom assembly that dictate the primary tensile force applied during the second stage pull. 3. The Two-Stage Pulling Mechanism Explained

2-Stage Pull: Heats and elongates the glass slightly during the first step, pauses to let it cool/reset, and then executes a final high-heat pull. Essential for ultra-fine patch-clamp pipettes or intracellular electrodes. narishige pc10 manual new

Located on the lower frame; controls incoming power to the primary transformers.

: Do not loosen internal structure clamps or attempt to disassemble for repair. 3. Operational Procedures Initial Setup

has improved digital displays, the core mechanics of the PC-10 remain the standard for reliable, manual pipette production. NARISHIGE GROUP Quick Setup Guide 1. Control Panel Overview Power Switch: Located on the front; the Heater Level display illuminates when on. Mode-Selector Knob: Single-stage pulling (one continuous pull).

Ensure the vertical tracks sit perfectly perpendicular to the benchtop. Controls the electrical current sent to the platinum

Check that the platinum heating element is centered. The glass capillary must pass directly through the middle of the coil loop without touching the metal walls.

Controls the electrical current delivered to the platinum coil. Higher numbers increase the temperature.

Realign the platinum coil so the glass capillary passes directly through its true center. Air drafts or dirty sliding rails.

A clear cover protects the pulling area from external air currents, which can otherwise cause inconsistent cooling and tip deformation. The capillary will pull down to a thin, tapered state

Before first-time use, wipe the vertical guide shaft with a lint-free cloth moistened with a small amount of light machine oil or alcohol to ensure friction-free travel.

Variable weights (sub-weights can be added or removed from the slider).

The updated Narishige PC-10 manual provides a comprehensive guide to the operation, technical specifications, and safety precautions of the device. The new features and instructions outlined in the manual will help researchers to optimize their use of the device and achieve accurate and reliable results.

This guide serves as a manual for new users, detailing its features, setup, and optimization strategies to ensure your research stays precise.

Getting the perfect pipette often requires optimizing the parameters, especially if you are using new glass or a new machine.